Pular para o conteúdo principal
GE Healthcare · MRExciteSigna

Unable to login to SCP/AGP/APS

Symptom

Unable to login to SCP/AGP/APS FE and/or OLE is attempting to login to the SCP/AGP/APS either using the "mgd_term" command or individually using the "CU SCP" or "CU AGP" or "CU APS" commands but the system won't let him/her/them do it. Instead they see two error messages for each attempt in each window, as follows: "Resource temporarily unavailable", and "All matching ports in use".

Error message

Resource temporarily unavailable Error in command window trying to log into SCP/AGP/APS. See Long Description. [insite@WSMI-MR1 ~]$ cu scp cu: open (/dev/ttyd043): Resource temporarily unavailable cu: scp: All matching ports in use [insite@WSMI-MR1 ~]$ ---------------------------------- [insite@WSMI-MR1 ~]$ cu aps cu: open (/dev/ttyd041): Resource temporarily unavailable cu: aps: All matching ports in use [insite@WSMI-MR1 ~]$

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2379186 Term Server
Area
MR

Full cause & step-by-step resolution