GE Healthcare · MRExcite22426642227275
Scan aborted due to arrythmia triggers caused by an irregular heart palpitation produced by the patient
Symptom
ECG and Cardiac Gating have no Signal Cannot change from peripheral gating to ECG gating (any lead), no signal. ECG and Cardiac Gating have no Signal
Error message
2242664 Unable to Find a Valid ECG Trigger for 10 Secs. Error: 2242664 Host: lx-mr Process: NSP File: AGP:wait for valid gate(): No ECG Line: 0 Timed out. Unable to find a valid ECG trigger for 10 secs. 2227275 Scan aborted due to 21 arrythmia triggers Thu Dec 29 08:43:24 2005 Error: 2227275 Host: HSC_MR01 Process: NSP File: HPC_PORT_ID:psd scan entry point Line: 0 Scan aborted due to 21 arrythmia triggers. This may have been caused by changed cardiac bpm while scanning. Please represcribe with an updated bpm, retry acquisition.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.