Pular para o conteúdo principal
GE Healthcare · MRExcite2255482

Unable to download PROPELLER scan prescription.

Symptom

Propeller Scan Option Will Not Run or Runs Very Slow propeller option will not function

Error message

2255482 Server Name: DP7 Unknown Server error DPServiceManager.cpp occurred See Long Description Error: 2255482 Host: lx-mr Process: LxMxConsumer File: DPServiceManager.cpp Line: 423 Server Name: DP7 Unknown Server error DPServiceManager.cpp occurred. -------------------------------------- Comments from engineering on this error message: Each Mercury-AP is actually 8 separate processors. "DP7" refers to the process that runs on the seventh processor of the AP, which is on the slave AP board. The error is reported probably because of a bad switch (8-port ethernet switch in the System cabinet) or a bad AP.

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2325947-12 AP Reflex Minerva II 400B Master Board - M3000RJ For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. 2325947-13 AP Reflex Minerva II 400B Board Slave Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. GE part-number Mercury part-number Description M3000RJ Reflex 400B 901-07014 System bundle consisting of 1 900-07071, 1 900-07072 2325947-12 900-07072 Minerva-2 motherboard configured as master 2325947-13 900-07071 Minerva-2 motherboard configured as slave GE part-number Mercury part-number Description M3000RJ Reflex 400B 901-07014 System bundle consisting of 1 900-07071, 1 900-07072 2325947-12 900-07072 Minerva-2 motherboard configured as master 2325947-13 900-07071 Minerva-2 motherboard configured as slave
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues