GE Healthcare · MRExcite2255482
Unable to download PROPELLER scan prescription.
Symptom
Propeller Scan Option Will Not Run or Runs Very Slow propeller option will not function
Error message
2255482 Server Name: DP7 Unknown Server error DPServiceManager.cpp occurred See Long Description Error: 2255482 Host: lx-mr Process: LxMxConsumer File: DPServiceManager.cpp Line: 423 Server Name: DP7 Unknown Server error DPServiceManager.cpp occurred. -------------------------------------- Comments from engineering on this error message: Each Mercury-AP is actually 8 separate processors. "DP7" refers to the process that runs on the seventh processor of the AP, which is on the slave AP board. The error is reported probably because of a bad switch (8-port ethernet switch in the System cabinet) or a bad AP.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2325947-12 AP Reflex Minerva II 400B Master Board - M3000RJ For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. For Excite 2 Systems. Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. 2325947-13 AP Reflex Minerva II 400B Board Slave Checkout Resolution 'Reflex AP Troubleshooting Tips" before ordering. GE part-number Mercury part-number Description M3000RJ Reflex 400B 901-07014 System bundle consisting of 1 900-07071, 1 900-07072 2325947-12 900-07072 Minerva-2 motherboard configured as master 2325947-13 900-07071 Minerva-2 motherboard configured as slave GE part-number Mercury part-number Description M3000RJ Reflex 400B 901-07014 System bundle consisting of 1 900-07071, 1 900-07072 2325947-12 900-07072 Minerva-2 motherboard configured as master 2325947-13 900-07071 Minerva-2 motherboard configured as slave
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.