GE Healthcare · MRExcite22523052252293
Cannot Scan with Head T/R Driver Error Due to SRFD2 RF Amplifier Flexible Extension Cable Open
Symptom
Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.
Error message
2252305 TR Driver Fault Detected Open Circuit Source of Fault is Body TR. Thu Aug 5 11:00:38 2004 Error: 2252305 Host: fmri3te Process: NSP File: SCP:DCB Line: 1197 TR Driver Fault Detected - Open Circuit. Source of Fault is Body TR. Value: 16 mA Limit: 2400 mA-> 2252293 The Driver Module Has Detected a Fault. Wed Apr 21 20:00:24 2004 Error: 2252293 Host: lx-mr Process: NSP File: SCP:DCB Line: 1986 The Driver Module Has Detected a Fault.^A Details of the Fault can be found in other error messages.->
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5116213-2 SRFD2 RF Amplifier Flexible Extension Cable 4 Foot cable coming off the rear of the amplifier and connects to the heliax cable.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.