Will not Scan with Table not at Scan Plane Errors Due to broken wire on home flag

Symptom

Will not Scan with Table not at Scan Plane Errors Intermittently the table will not be at scan location and numbers on gantry will not change. Advancing the table does not update the encoder. Also no LEDs are lit on the keyboard. Intermittently the table will not be at scan location and numbers on gantry will not change. Advancing the table does not update the encoder.

Error message

2224666 Auto Prescan failed The table is not at scan plane Wed Oct 19 18:25:22 2005 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, The table is not at scan plane.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.