System fails to boot Application Software.
Symptom
System fails to boot Application Software. Can't ping anything in the MGD. Excite HD upgrade. First boot fails. Cant ping the Term Server or any of the boards in the MGD.
Error message
2229307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Value 0x5555 not found at address 0xC41000A8 after 20000 msec.-> 2249372 Communication problem with the host. Remote command execution (rcmd) failed. See long description for complete error. Fri Apr 1 16:52:38 2005 Error: 2249372 Host: lx-mr Process: diagCoord File: mgdPosixSync.c Line: 1215 Task ID: 0x3ff00c0 Time: 1706269 Object: APS Communication problem with the host. Remote command execution (rcmd) failed. Host: host; RSH Port: 514; Local User Name: genesis Remote User Name: genesis; Remote Command: getHostTime; fd2p pointer: 0-> 2223849 AGP, SCP, APS EPROM power-up boot failed 2249424 Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. see long description 1094564028 0 1 Tue Sep 7 13:33:48 2004 2249424 2 CHMMR diagCoord mgdPosixSync.c 1081 Task ID: 0x3ff00c0 Time: 32822 Object: APS Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. Host: host; SNTP Timeout: 100; Errno: 0x710005-> EN 298 SR 298 200002352 A HOST_SHUTDOWN was initiated by the death of RTPSIServer See Long Descripti
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.