GE Healthcare · MRExcite2229307200002249372170626922238492249424
System fails to boot Application Software.
Symptom
System fails to boot Application Software. Can't ping anything in the MGD. Excite HD upgrade. First boot fails. Cant ping the Term Server or any of the boards in the MGD.
Error message
2229307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Value 0x5555 not found at address 0xC41000A8 after 20000 msec.-> 2249372 Communication problem with the host. Remote command execution (rcmd) failed. See long description for complete error. Fri Apr 1 16:52:38 2005 Error: 2249372 Host: lx-mr Process: diagCoord File: mgdPosixSync.c Line: 1215 Task ID: 0x3ff00c0 Time: 1706269 Object: APS Communication problem with the host. Remote command execution (rcmd) failed. Host: host; RSH Port: 514; Local User Name: genesis Remote User Name: genesis; Remote Command: getHostTime; fd2p pointer: 0-> 2223849 AGP, SCP, APS EPROM power-up boot failed 2249424 Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. see long description 1094564028 0 1 Tue Sep 7 13:33:48 2004 2249424 2 CHMMR diagCoord mgdPosixSync.c 1081 Task ID: 0x3ff00c0 Time: 32822 Object: APS Communication problem with the host. Simple Network Time Protocol (SNTP) request failed. Host: host; SNTP Timeout: 100; Errno: 0x710005-> EN 298 SR 298 200002352 A HOST_SHUTDOWN was initiated by the death of RTPSIServer See Long Descripti
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.