GE Healthcare · MRExcite2248199
After initial powerup, 20Mhz clock failed on the RF-DIF board in the RRF Chassis
Symptom
20Mhz clock failed on the RF-DIF board in the RRF Chassis Happened after initial power up.
Error message
2248199 APS IrfDriver RRF link tx not ready error 7352 Thu May 19 10:38:33 2005 Error: 2248199 Host: pzmrov Process: NSP File: APS:IrfDriver::irfInit Line: 5665 IrfDriver RRF link tx not ready error 7352 Please perform a TPS Reset. 2248199 SCP IrfDriver RRF link tx not ready error 3256 Thu MayWed May 18 09:03:28 2005 Error: 2248199 Host: pzmrov Process: NSP File: SCP:IrfDriver::irfInit Line: 5739 IrfDriver RRF link tx not ready error 3256 Please perform a TPS Reset.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2308900 RF MODULE W/POWER SUPPLIES
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.