GE Healthcare · MRExcite2228179224859322483051019722092842218537
Scan Paused Due To SCP Board in the MGD
Symptom
HP Computer Hang
Error message
2228179 TPS detected NSP (host) timeout Tue Sep 20 14:17:14 2005 Error: 2228179 Host: lx-mr Process: NSP File: APS:hs watchdog check Line: 10 TPS detected NSP (host) timeout and terminated any scan in progress. 2248593 SCP Waveform Buffer Overflow 2248593 Host: lx-mr Process: NSP File: SCP:OpWorkspaceHostHWO Line: 548 SCP waveform buffer overflow. A TPS reset may be necessary. 2248305 It may be necessary to perform a TPS Reset. An SCT packet send failed. Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Internal Software Error. 2248305 lx-mr NSP SCP:scp_send_pkt 10 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 Internal Software Error. It may be necessary to perform a TPS Reset. An SCT packet send failed. Packet opcode: 10197 Status: 2209284 2218537 Scan stopped, Got Recon-Scan Stop packet With Error Message Key Fri Apr 16 10:11:34 2004 Error: 2218537 Host: LX1 Process: scn File: MsaScanMgr.m Line: 2641 Scan stopped, Got Recon-Scan Stop packet with error message key - 2248
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- 2294300-12 SCP2 Board See Long Description 2294300-12 SCP2 Board. FEMC Parts Comments states as follows: If no stock then 2294300-19 is a suitable substitute. 212002950 February 8, 2010.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.