GE Healthcare · MRExcite2248840
No receive signal when scanning body or surface coils.
Symptom
Unable to scan with body or Surface Coils with RF Errors No receive signal when scanning body or surface coils. Head scans ok.
Error message
2248840 Analogic RF Amplifier 2 Voltage Standing Wave Ratio (VSWR) Fault. Wed Aug 10 12:48:50 2005 Error: 2248840 Host: lx-mr Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2492 Analogic RF Amplifier 2 Voltage Standing Wave Ratio (VSWR) Fault. Analogic RF Amplifier 2 reflected power exceeds 2250W. Possible causes: disconnected or defective output cable from RF Amplifier 2 to RF Interface (RFI), defective RFI or RF Amp 2.->
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Component
- SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.