No receive signal when scanning body or surface coils.

Symptom

Unable to scan with body or Surface Coils with RF Errors No receive signal when scanning body or surface coils. Head scans ok.

Error message

2248840 Analogic RF Amplifier 2 Voltage Standing Wave Ratio (VSWR) Fault. Wed Aug 10 12:48:50 2005 Error: 2248840 Host: lx-mr Process: NSP File: SCP:../RfNbMdsHWO.cpp Line: 2492 Analogic RF Amplifier 2 Voltage Standing Wave Ratio (VSWR) Fault. Analogic RF Amplifier 2 reflected power exceeds 2250W. Possible causes: disconnected or defective output cable from RF Amplifier 2 to RF Interface (RFI), defective RFI or RF Amp 2.->

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573 See comments in Resolution Concept for second opinion requirements before ordering this part. SecondOpinion SRFD2 1.5T Solid State RF Amplifier 2351573
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.