Two PDU cables, 2371514 (run# 968) and 2371515 (run# 1081) are mislabeled

Symptom

Two PDU cables, 2371514 (run# 968) and 2371515 (run# 1081) are mislabeled on the PDU side. See Long Description Subject: Mis-labeling of PDU cables for HFO4 and HFO5 SMILe 08/16/2005 Effectivity: HFO4 and HFO5 Problem Description: Two of PDU cables which are both 2371514(run# 968) and 2371515(run# 1081) have a mis-labeling. At the end of PDU side, a cable label of 2371514(run# 968) says "MR3-PD1 4,NEU,5,GND,6" but those should be "MR3-PD1 16,NEU,17,GND,18". A cable label for 2371515(run# 1081) says "MR3-PD1 18,NEU,19,GND" but those should be "MR3-PD1 14,NEU,15,GND".

Applies to

Product
Open Speed with EXCITE For HFO4 and HFO5 only
Software
OPENMR4-HFO4 Software version for HFO4 system, which is OpenSpeed with Excite Technology