GE Healthcare · MRExcite
AP Functional Check in Diagnostics Reports a configuration failure for Reflex 200B board configuration
Symptom
AP Functional Check in Diagnostics reports a configuration failure for Reflex 200B board configuration See Attachment for the description and a snapshot of the problem.
Error message
Warning: This is not a standard AP configuration. There should be 1708 meg of AP BAM memory, you have 844 meg. One or more of the AP boards may be partially defective Warning: This is not a standard AP configuration. There should be 8 AP CPUs, you have 4. One or more of the AP may be partially defective.
Applies to
- Models
- Excite
- Product
- Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
- Software
- 11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.