Gradcal has distortion on X image and over range after HD upgrade Due to HFA Control Board

Symptom

Gradcal has distortion on X image and over range after HD upgrade Gradcal has distortion on X image and over range after HD upgrade. All other planes are fine.

Error message

2245300 GRADIENT DRIVER TESTS - Gradient Fidelity Test failed on the % Axis 2245300 GRADIENT DRIVER TESTS - Gradient Fidelity Test failed on the Z Axis. EPI Asymmetry measured -344.471 uAs at Shot 0, View 2, Sample 51. ->

Applies to

Models
Excite, Signa
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR