Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor.
Symptom
Xray aborted, stopped- Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor. This is any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). Xray aborted - Restart Xray; if problem persists, call Service is displayed on the Innova Image monitor This is caused by any defect in a vascular system that prevents it from making X-rays for at least one request (Fluoro or Record). Sometimes, a reset system may be needed to recover X-rays (case IH error follows abort error). errors: 600104 IN2K: Error 4F76
Error message
IN2K: 1800018 ERR SIB IP BOARDS LINKS IP1 (card 0) -IP2 (card 1) link failure detected 1800018 ERR SIB IP BOARDS LINKS IH XRExpSens DSP init Component(s) RTAC SIB C80 Boards links 1 0 [DSP automaticNaming] IP1 (card 0) -IP2 (card 1) link failure detected dsp driver.cpp 635 Xray not available after changing the C80 board: 1800018 Xray not available after changing the C80 board. Error 1800018 Error Text ERR_SIB_IP_BOARDS_LINKS Debug Text [DSP_automaticNaming] SIB-IP1 (card 0) link failure detected
Applies to
- Models
- Innova
- Product
- Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
- Component
- Digital Leader (DL) Digital Leader Digital Leader (DL)
- Area
- X-ray Vascular
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray Vascular issues
- The Worklist does not refresh on the system when performing a query.
- Innova System boot failure
- INX1X; Message Limit switch engaged; LP4 Lateral Detector lift will not move after going to max SID using the Emergency Backout button on the lift cover because the lift hit maximum (Max) SID limit sw…
- Firewall is broadcasting DHCP addresses
- Cooling Failure; Shut System down and call Service is displayed on the Innova Image monitor
- No gantry motion is reported by the customer.