The Customer complains of a Water or a fluid leak on the system.

Symptom

The Customer complains of a Water or a fluid leak on the system. The customer indicates there is water leaking from the closed loop cooling system. Fluid is coming out of the tube arm assembly or in the technical room. If zero stock on 2184070-2 RP ROTATING JOINT then order 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next. The Customer complains of a Water or a fluid leak on the system. The customer indicates there is water leaking from the closed loop cooling system. Fluid is coming out of the tube arm assembly or in the technical room. 2184070-2 RP ROTATING JOINT then order 2129844-2 RIGID WATER COOLING ASSEMBLY and it contain the joint part. Also replace the two O-rings with new or they will leak next

Applies to

Models
Advantx, Innova
Product
Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems Advantx E Biplane Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane)
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required