A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector.
Symptom
A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. See long description for pop up message picture. A pink Attention box pops up during Fast Cal saying there is a problem with the DAS/Detector. Scanning will NOT be disabled. Image Quality will be compromised, call GE SERVICE. After clicking OK Fast Cal completes. ATTENTION: The system detected a problem with the DAS/Detector sub-system. Image Quality will be compromised. Scanning will NOT be disabled, but it is critical that the user call GE SERVICE to rectify the problem. Press 'OK" to proceed at the risk of Image Artifacts
Error message
Error: 260132753 Digital Interface Board(s) Encountered An Error During ASIC Linearization See Long Description for complete error Mon Oct 31 07:18:07 2005 Error: 260132753 Host: orp Process: orp_subsys File: Ifb.cxx Line: 504 Mon Oct 31 07:18:06 2005 Host : Das Ermes # : 260132753 Exception Class : Secondary Severity : Sec/Soft File : Ifb.cxx Line# : 504 Function : No System Function Reported Scan Type : Axial Scan: 50043/5/7 Digital Interface Board(s) Encountered An Error During ASIC Linearization DIFB : 4 Chan : 1430 Stage : 2 This error format was introduced in 07MW18.4_SP2-26.V40_H_H64_G_GTL with SP2 release . We no-longer need the decoder since this error report the FRDM as A/D 41 . Wed Jan 7 08:40:45 2009 Error: 260132753 Host: orp Process: orp_subsys File: Ifb.cxx Line: 890 Wed Jan 7 08:40:44 2009 Host : Das Ermes # : 260132753 Exception Class : Abort Severity : Secondary File : Ifb.cxx Line# : 890 Function : No System Function Reported Scan Type : Axial Scan: 50018/1/3 Digital Interface Board(s) Encountered An Error During ASIC Linearization DIFB : 21 A/D Board: 41 Chan : 3369 Stage : 3 Coeff : 22 Other Examples: From software releases prior to 2009, the following error fo
Applies to
- Models
- Optima, VCT, BrightSpeed
- Product
- Optima CT660 10HW02.10.H40_P2_HS64_G_GMV 10HW33.8 H40_P2_HS64_G_GMV Optima CT660 Optima CT660 is a new design 64 slice scanner based on VCT and BrightSpeed technology. This is the first CT scanner as new Optima brand. check_config Product: Optima CT660 sw_version: 10HW02.10 or 10HW33.8 sw_hw_string: 10HW02.10.H40_P2_HS64_G_GMV or 10HW33.8_SP3-0-1.H40_P2_HS64_G_GMV host_id: 1943170864 tube_type: 19 gantry_type: GANTRY_HP_40 No. of processors: 8 Processor speed: 2527 memory: 24576 hinv CPU: GenuineIntel Intel(R) Xeon(R) CPU E5540 @ 2.53GHz 8 2527 MHz x86_64 Processors Secondary cache: 8192 KB FPU: GenuineIntel Intel(R) Xeon(R) CPU E5540 @ 2.53GHz Floating Point Main memory size: 24576 Mbytes SCSI controller 10: Version usb-storage Disk drive: unit 00 on SCSI controller 10, Lun 00 SCSI controller 9: Version usb-storage Disk drive: unit 00 on SCSI controller 9, Lun 00 SCSI controller 0: Version mptsas Disk drive: unit 00 on SCSI controller 0, Lun 00 Disk drive: unit 01 on SCSI controller 0, Lun 00 16550A UART serial port: 000003F8 Graphics board: VGA Ethernet interface configured: eth3 Ethernet interface configured: eth0 Ethernet interface configured: eth1 Dual Channel Display Table weight limit: GT1700V Table The Lite table ( currently for Japan Market) 450 Pounds (205 kg) Table: The normal table is 500 Pounds (227 kg) Table: X-Ray Tube Modified Performix D3187T Green Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 5202201 FRDM Single FL Outer Module for HDAS_SATURN_64 See Long Description Part #: 5202201 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 3.31 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: V-RES FIELD REPLACEABLE MODULE KIT - FL SLOTS FRU: Yes Status: Orderable Parts Offered as: Current : 5202201 Harvested : Not Available Exchange : Not Available Comments: Per Titan RSPL, 212008898, 1/15/2010.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.