GE Healthcare · MROptimaExcite2254415
RF Amplifier in service, not product mode Due to UPM NB IF Board.
Symptom
RF Amplifier in service, not product mode RF Amplifier in service, not product mode.
Error message
2254415 Attention: RF Amplifier in service, not product mode 2254415 NSP SCP:../rf_debug.c 443 Exception Class: Unknown Severity: Unknown Function: RF: RF Transmit PSD: Unknown Coil: Unknown Scan: -1/-1 Attention: RF Amplifier in service, not product mode. The RF Amplifiers are being controlled via the serial terminal. Reset the RF Amplifier Interface or Reset TPS to return control to the system.
Applies to
- Models
- Optima, Excite
- Product
- Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
- Software
- DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
- Component
- 2391146 UPM NB AMP IF BOARD, also known as ASC RF Interface Board.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.