GE Healthcare · MRExcite222467222278342224666
Data Acquisition Errors Due to STIF Board.
Symptom
Auto Prescan Failed with Data Acquisition Time Out
Error message
2224672 Auto Prescan failed, Data Acquisition time out 2227834 A Data Acquisition Error Occurred 2227834 1 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. AP board level diagnostics, and Image Transfer (IT) board level diagnostics. Multiple iterations may be required before the error happens. 00000138H EN 314 ER 395 2227834 1 ex-mr NSP APS:da_dacq_task 21 A data acquisition error occurred. Run datapath diagnostics. 00000068H 2224666 Auto Prescan failed, Auto Prescan failed, possible data acquisition failure. 2224666 mmp-mr01 psc PscMainControl.cc 1311 Auto Prescan failed, possible data acquisition failure.
Applies to
- Models
- Excite
- Product
- 3T Excite G3 Software ONLY
- Software
- G3 For 3T Excite only
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.