Unable to scan with CAN errors due to shim power supply.
Symptom
Unable to Scan with CAN Network Errors See long description MR SMILe: March 17, 2005 Certain conditions can infrequently occur with devices on the CAN Link that will result in CAN Link errors and the loss of communication with these devices. Devices such as Driver Module, RF-DIF Board, GP3 Module, ASC (UPM) components, HOS, and Patient Handling Power Supply can be affected. CAN Link errors in the log and general troubleshooting oftentimes implicates one device on the CAN Link and results in it's replacement. Changing the part usually fixes the problem, but sometimes it does not. Another device on the CAN Link is later determined to be disrupting communications.
Error message
2252232 SCP detected a Node Reset on the CAN network 2252232 Host: EX3T Process: NSP File: SCP:CanLinkHWO Line: 617 SCP detected a Node Reset on the CAN network. Reset Node is Master RRF-DIF. Thu Aug 16 02:53:42 2007 Error: 2244591 Host: EX3T Process: NSP File: SCP:Grad Line: 3011 The GP is reporting a SGA Power Supply OverCurrent Fault.- :
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 5106397 5 Channel Shim Power Supply
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.