Pular para o conteúdo principal
GE Healthcare · MRExcite22522322244591

Unable to scan with CAN errors due to shim power supply.

Symptom

Unable to Scan with CAN Network Errors See long description MR SMILe: March 17, 2005 Certain conditions can infrequently occur with devices on the CAN Link that will result in CAN Link errors and the loss of communication with these devices. Devices such as Driver Module, RF-DIF Board, GP3 Module, ASC (UPM) components, HOS, and Patient Handling Power Supply can be affected. CAN Link errors in the log and general troubleshooting oftentimes implicates one device on the CAN Link and results in it's replacement. Changing the part usually fixes the problem, but sometimes it does not. Another device on the CAN Link is later determined to be disrupting communications.

Error message

2252232 SCP detected a Node Reset on the CAN network 2252232 Host: EX3T Process: NSP File: SCP:CanLinkHWO Line: 617 SCP detected a Node Reset on the CAN network. Reset Node is Master RRF-DIF. Thu Aug 16 02:53:42 2007 Error: 2244591 Host: EX3T Process: NSP File: SCP:Grad Line: 3011 The GP is reporting a SGA Power Supply OverCurrent Fault.- :

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Component
5106397 5 Channel Shim Power Supply
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.