GE Healthcare · MRExcite225385025412572549307254470722278342253613
Data Acquisition errors during scan.
Data Acquisition Error
Symptom
Data Acquisition errors during scan. Data acquisition error occurred. Run MGD datapath diagnostics, AP board level diagnostics, and Image Transfer (IT) board level diagnostics. Multiple iterations may be required before the error happens. 000000ceH Also gradient hammer tests fail.
Error message
2253850 Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:22:58 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2541257 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid X axis data value via the gradient data link. Data Exp = 0007ffffH or 00080000H Data Rd = 00080010H Check the gradient data and clock fiber optic cable connections 2253850 Gradient Link Hammer test failed. The GP received an invalid Z axis data value via the gradient data link. See long description for entire error. Sun Jun 19 00:24:18 2005 Error: 2253850 Host: lx-mr Process: diagCoord File: grad_hammer.c Line: 286 Diagnostic: Grad Hammer Iteration: 8 of infinite Action: 1 Task ID: 0x3d5c6c8 Time: 2549307 Object: SCP Test Task Gradient Link Hammer test failed. The GP received an invalid Z axis data value via the gradient data link. Data Exp = 000aaaaaH or 00055555H Data Rd = 000aaabaH Check the gradient data and clock fiber optic cable connections 2253850 Gr
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.