GE Healthcare · MRExcite2248121
Image ghost is intermittent and must power off for several hours, it will scan normal.
Symptom
MRIQ Intermittent Ghosting
Error message
2248121 APS failed to send command message to DRF 40 DRF might have crashed because of other errors. Please perform a TPS Reset.-> See Long Description 2248121 MR3T NSP APS:DrfDriver::sendCmdMsg 1068 APS failed to send command message to DRF 40 DRF might have crashed because of other errors. Please perform a TPS Reset.-> Extended Error: The reset should clear the hardware. If not, run the MGD Board Level and Datapath diagnostics. Also, try reseating the boards listed below. Possible cause(s): Software,
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Component
- 2274196 IRF I/O Board See Long Description. Per FEMC part information, Description of part number 2274196 is an INTERFACE and REMOTE RF TRANSITION BOARD. Part Comments as of 2/8/2008 are as follows; FLS - INTERFACE & REMOTE RF TRANSITION ADDED PER RSPL. USED ON PHOENIX MGD. MSA8429 05/14/02 ROUTE TO MR FOR CONSUMABLE COLLECTION PER B. HAHLEN, 7-11-02, A. MASCH THIS PART WAS ADDED AS PART OFTHE EXCITE NPI FOR MR. 9/01/02 MS24445 BH FLS - FE ASSIST 08OCT2002 (MS2045) ADDED TO MR DISTRIBUTION LIST FOR PARTS TRACKING.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.