Unable to Boot Applications Software with TPS Download Failure due to SCP2 Board in MGD Chassis.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 2223849 SCP EPROM power-up boot failed (See attachment for extended error message) 2223849 rmny NSP TardisReset.cc 1841 Exception Class: Unknown Function: Host: NSP SCP EPROM power-up boot failed. Communication Timed-out from SCP board. Verify that MCP820 boards are installed and seated properly for APS and AGP.-> 2249307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Error seen with MGD Diags: 1081934830 0 1 Wed Apr 14 09:27:10 2004 2249307 2 CARD diagCoord pearlMap.c 1806 Task ID: 0x0 Time: 0 Object: SCP Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Value 0xAAAA not found at address 0xC03010A8 after 20000 msec.->
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Component
- 2294300-12 SCP2 Board See Long Description 2294300-12 SCP2 Board. FEMC Parts Comments states as follows: If no stock then 2294300-19 is a suitable substitute. 212002950 February 8, 2010.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.