GE Healthcare · MRSignaExciteDiscoveryOptima2228116225801322552032244597
Software corruption due to host date and/or time changed with GE scanner application software running, and not immediate…
Software corruption due to host date and/or time changed with GE scanner application software running, and not immediately rebooting.
Symptom
DWI images coming out as blank gray image, with Calibration Data file errors DWI images coming out as blank gray image, with Calibration Data file errors.
Error message
2228116 Data may be corrupted. Image quality may suffer. See Long Description Error: 2228116 PSD: Spin Echo Coil: 8NVARRAY A Scan: 3794/10 Data may be corrupted. Image quality may suffer. 2258013 Data processing error Component CalibrationDataService has received an invalid event 2258013 Host: MR3T Process: LxMxConsumer File: CalibrationDataService.cpp Line: 1807 Server Name: DP Data processing error: Component CalibrationDataService has received an invalid event. 2255203 File /export/home/signa/recon_calib/4560_432_0_Calib.dat can not be opened 2255203 Host: MR3T Process: LxMxConsumer File: extensions/ace/GeOS.cpp Line: 113 Server Name: DP File /export/home/signa/recon_calib/4560_432_0_Calib.dat can not be opened 4003 Error while formatting Ermes message #2244597 - illegal character in arguments. See Long Description 4003MR01 NSPdoBformat.c 1108Error while formatting Ermes message #2244597 - illegal character in arguments.
Applies to
- Models
- Signa, Excite, Discovery, Optima
- Product
- Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.