GE Healthcare · MROvationExciteSigna2225541
Unable to Scan with Dynamic Disable Faults due to SSM.
Symptom
Unable to Scan with Dynamic Disable Faults Unable to Scan with Dynamic Disable Faults.
Error message
2225541 Dynamic disable fault, Rear of Coil open. Dynamic Disable bias fault on back drivers. Wed Dec 1 09:19:34 2004 Error: 2225541 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_tr_mon Line: 6160 Dynamic disable fault, Rear of Coil open. Dynamic Disable bias fault on back drivers. Possibly a disconnected cable between the driver board and the dynamic disable switch boxes, or an open diode in dynamic disable switch boxes on BACK of BRM.
Applies to
- Models
- Ovation, Excite, Signa
- Product
- Ovation/MFO
- Software
- MFO3 Applies to Ovation3 with HP4000 computer running Linux OS.
- Component
- 2160200-45 SSM assy for Ovation/OpenSpeed
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- CWA ID100060 Unable To Film.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.