GE Healthcare · MRExcite2224666

Table Not at Scan Plane Due to Longitudinal Encoder.

Symptom

Auto Prescan Failure when table is not at scan plane. see long description When the system requires the user to move the table, the Move to Scan button on the operator’s console will blink green. After the user presses this button, the table does move, but depending on the timing, an error message appears stating “Unable to start prescan! Table not at scan plane”. Workaround : After the table has moved to the scan plane, push the Scan button again and the sequence will start. See page 2 of service note DOC0765967 "MR450 and MR750 DVIB1 Existing Issues, Workarounds, and Plans". Service notes can be found at http://3.87.118.28/optec3/common/fmisn/sn.htm Permanent Fix: 20.1_IB2 FMI 60787

Error message

2224666 Auto Prescan failed The table is not at scan plane Wed Oct 19 18:25:22 2005 Error: 2224666 Host: lx-mr Process: psc File: PscMainControl.cc Line: 1322 Auto Prescan failed, The table is not at scan plane.

Applies to

Models
Excite
Product
3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
Software
14.0 M5 Excite HDx HDi Platforms only.
Component
5224109 Rotary Table Position Encoder 5224109 Rotary Table Position Encoder
Area
MR

Full cause & step-by-step resolution