GE Healthcare · X-ray VascularInnova3602085380033136020163602087

Tube Spits in Record, not in Fluoro or X220:Innova X100 Monoplane:Replaced Jedi Ingrid tank

Symptom

INX100 : Tube Spits in Record, not in Fluoro or X220 When being in application in fluoro mode it was OK but in the first exposure in Cardiac, I got the message X-RAY UNAVAILABLE RESET FOOT SWITCH AND TRY AGAIN. The image displayed was bad and another message THE IMAGE IS DEGRADED TRY AGAIN appeared on the console and IRD. Issue could not be reproduced in Calibration unit X220.

Error message

KV REGULATION SPITS. 3602085, error Text = Jedi: too many tube spits, debug Text += jedi-error=316h, Spit Max error, data related to error: 8, recovery Class = AR, major Function /= XRTechMgr KV REGULATION SPITS. System impact: X-ray unavailable. Displayed message : "Xray aborted - Restart Xray; if problem persists, call Service". May be preceded by error 3800331 Bad actual techniques recieved in aquisition, and followed by 3602016 Degraded Image Quality

3602087 Tube spits error details kV reg on steady state: 7 3602087 Jedi: Tube spits error details DO XRTechMgr control JEDI tube spits details 1 0 Anode: 0 Cathode: 0 Anode+Cathode: 0 kV reg on steady state: 7 kV reg on kV rise: 0 XRTechMgr.cpp 2272 38 2007

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare X-ray Vascular issues