GE Healthcare · MRExciteSigna2252312
Unable to Scan with Dynamic Disable Faults due to shorted RF body coil.
Symptom
Unable to Scan with Dynamic Disable Faults Unable to Scan with Dynamic Disable Faults.
Error message
2252312 - DD Driver Fault Detected - High Voltage Fault. Error: 2252312 Host: mr3t Process: NSP File: SCP:DCB Line: 1884 DD Driver Fault Detected - High Voltage Fault. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value = 213 V Limit: 425 V->
Applies to
- Models
- Excite, Signa
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- SecondOpinion 3T Short RF Body Coil with FRU case 2208999-7 and 5133040 is the body coil alone See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR