Pular para o conteúdo principal
GE Healthcare · CTBrightSpeedVCT200109012200502005

BrightSpeed HPower, Delight, Velocity or MERI HPower

Symptom

Image saving error; database restoration in progress

Error message

Error: 200109012 Image Allocation Failed. Cannot build scan request. See Long Description for complete error Wed Mar 2 18:39:33 2005 Error: 200109012 Host: ct01 Process: scanRx File: UIRxToScanner.cxx Line: 1307 Function: Data Acquisition : OC Processing Image Allocation Failed. Cannot build scan request.

Error: 200502005 Exam not found in image management, cannot reallocate image space. See LD for complete error Mon Mar 26 12:51:14 2007 Error: 200502005 Host: roct Process: scanRx File: srx_sma_manage_image_space.c Line: 1059 Function: No system function reported ROCT 736/999/0 not found in image management, cannot reallocate image space.

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed HPower, Delight, Velocity or MERI HPower HP_P_G4_G_HPT Excel HP_P_G8_G_HPT Edge HP_P_G16_G_HPT Elite HP_P_G16_G_GMV Elite HP_P_P16_G_GMV Elite HP_P_M16_G_HPT BrightSpeed Method of remote software hardware identification check_config swhwinfo List of Supported Console Types : HP_P_G16_ G _HPT Note: G could be GRE, AIO or TIO - check with: cat /usr/g/config/INFO | grep CONSOLETYPE GOC4 AIO TIO List of softwares the product can run: 08BW20.1 07MW11.10 08BW44.1 ( Delight ) 09HW30.4 ( Velocity ) 09BW35.11 ( MERI - Pancake DAS ) FAQ's 1. All BrightSpeed Power Systems use the Performix tube D3186T HP_P_G4_G_HPT (Excel - H-Power Table) HP_P_G8_G_HPT (Edge -H-Power Table ) HP_P_G16_G_HPT (Elite -H-Power Table ) HP_P_G16_G_GMV (Elite -Global Table Mid Value - modified GT1700 no IMS ) Note: G could be GRE, AIO or TIO (check with: cat /usr/g/config/INFO | grep CONSOLETYPE ) BrightSpeed Elite Power Systems will only be sold as Velocity System starting in 2Q/2010. Delight Elite Power being phased out 1Q/2010 BrightSpeed Power 4 & 8 slice systems have been discontinued due to restructuring of the product lineup. Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5117866-51 Host Computer, CTPET HP8200 with FX1500 Graphics Card and 4GB Memory - replaces -24, -28, and -48 versions See Long Discription for Details Part #: 5117866-51 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CTPET HP8200 with FX1500 Graphics Card and 4GB Memory FRU: Yes Status: Orderable Comments: Per ECO 2042338, 5117866-51 is a substitute for 5117866-28 for Brightspeed systems. 212008898, 6/13/07. ** FE Assist ** For VCT systems must have software release 07MW18.4, p/n 5213049 (deployed by FMI 25394, p/n 5195244), and perform a LFC to support this FRU if system is not currently running sw rls 07MW18.4. or newer. For other CT systems , must order sw releases 07MW11.10, p/n 5220443 (deployed by FMI 25389, p/n 5191771)and perform a LFC to support this FRU if system is not currently running the sw release, or newer (9/12/07 212001261) 5115333 VCT GOC5 HOST HP 8000 computer New part number 5117866-28 Description: Host Computer, CT USA VCT XW8200 Part #: 5117866-28 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CTPET HP8200 with FX1400 Graphics Card and 4GB Memory FRU: Yes Status: Orderable Comments: Added per VCT RSPL, 040522 ms13452 This part is direct replacement for 5115333 . ms13452 Part #: 5117866-24 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 60.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: Host Computer, CT USA 4,8,16-slice XW8200 FRU: Yes Status: Orderable Comments: Added per pathfinder RSPL, this part is not replacement validated for xw8000. ms13452 *** 2ND OPINION NOTES *** This part has a High NDF ( No Defect Found ) rate. Attempt to reload OS and Application software before replacement. ( 212024490 - 12/20/2006 )
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.