Pular para o conteúdo principal
GE Healthcare · MRExciteSigna222466622540522254051

Will Not Scan With Power Monitor Faults Due To The SSM.

Symptom

Unable to scan with Power Monitor Faults

Error message

2224666 Auto Prescan Failed, RF Amplifier is Not Ready NOTE: When using this error to create new knowledge, please include any associated error codes. 2224666 is a generic error code which requires further information. Mon May 17 08:19:08 2004 Error: 2224666 Host: lx_mr Process: psc File: PscMainControl.cc Line: 1296 Auto Prescan failed, RF amplifier is not ready.

2254052 Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and Patient Weight Fri Aug 20 08:54:29 2004 Error: 2254052 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Power Monitor 2 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. Check patient weight entered. amplitude: 35 pulse width: 33 duty cycle: 09 Fault in head amplitude. If you need to confirm the source of the fault, reset the power monitor and repeat the scan

2254051 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan parameters and patient weight. See Long Description Fri Aug 20 08:33:11 2004 Error: 2254051 Host: lx-mr Process: NSP File: SCP:PmMdsHWO Line: 1019 Power Monitor 1 Fault. RF Transmit power has reached safety limit allowed for scan

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.