GE Healthcare · MRSignaExcite225440322611762252239222381722611802262850
TPS reset fails due to Driver Module.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2254403 Unblank cable not detected by RF Amplifier Interface. Make sure the unblank cable is connected in order to scan. See Long Description 2254403 Unblank cable not detected by RF Amplifier Interface. Make sure the unblank cable is connected in order to scan.-> 2261176 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error: An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. 2261176 tempmr NSP SCP:SriAutoConfig 796 SRI to 16 Channel Switch Communication Error:An error occurred reading the serial ID on the 16 Channel Switch. Please check the connection between the SRI and switch and make sure the switch has power. 2252239 Startup Timeout was detected on the Drivers Control Board 2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS. 2261180 SRI to IRFM1 Communication Error: An error occurred reading the se
Applies to
- Models
- Signa, Excite
- Product
- 3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
- Software
- 15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
- Component
- SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- Running grafidy3, and find there are spikes on the waveforms in the manual prescan display, both windows.
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- Fails to Scan with TORSO Coil.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- AE title has changed on destination host.