2252304 Multi Coil Bias Fault Due to Driver Module for Excite.

Symptom

Unable to Scan With Multi Coil Bias Faults Unable to scan with Multi Coil Bias Faults

Error message

2252304 Multi coil Bias Fault Receive Fault. Wed May 31 11:05:41 2006 Error: 2252304 Host: PHDGE3TMR Process: NSP File: SCP:DCB Line: 1143 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Spin Echo Coil: XXXXX Scan: 50334/1 Multicoil Bias Fault - Receive Fault. Try re-seating all coil connectors. MCD2 Channel 4 of 16, counting from 1 Value: 125 mA, Limit: 98 mA Mask of failed channels: 0xd8 MCD1 1-8 : A or C MCD1 9-16 : B MCD2 1-8 : B or Legacy MC MCD2 9-16 : C->

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution