The HFA Control Board was tripping the over temperature fault when a FIESTA-C scan was running. "Y-Axis Overtemperature…

The HFA Control Board was tripping the over temperature fault when a FIESTA-C scan was running. "Y-Axis Overtemperature Fault" Scanning FIESTA-C Protocol, Scan Aborts.

Symptom

"Y-Axis Overtemperature Fault" Scanning FIESTA-C Protocol, Scan Aborts Unable to complete a FIESTA-C scan. Sequence>TE=minimum, Flip Angle=75, Band Width=62.5, Freq Matrix=384, Phase Matrix=256, NEX=2 FOV=20, SliceThickness=1, Locs per Slab=32 The scan aborted with error code 2244579 "The GP is reporting a Y Axis OverTemperature Fault. System will wait 5 minutes before attempting recovery.->. Changed Rx to FIESTA by copying FIESTA-C sequence (except for NEX so as to maintain the same scan time) and had no problem. The problem only occurs with FIESTA-C Unable to complete a FIESTA-C scan. Sequence>TE=minimum, Flip Angle=75, Band Width=62.5, Freq Matrix=384, Phase Matrix=256, NEX=2 FOV=20, SliceThickness=1, Locs per Slab=32 The scan aborted with error code 2244579 "The GP is reporting a Y Axis OverTemperature Fault. System will wait 5 minutes before attempting recovery.->. Changed Rx to FIESTA by copying FIESTA-C sequence (except for NEX so as to maintain the same scan time) and had no problem. The problem only occurs with FIESTA-C

Error message

2244579 The GP is reporting a %x Axis OverTemperature Fault. System will wait 5 minutes before attempting recovery 2244579 The GP is reporting a %x Axis OverTemperature Fault. System will wait 5 minutes before attempting recovery.->

Applies to

Models
Signa, Excite
Product
Signa HDxT This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software. This product is for 1.5T Signa HDxT systems with 15.0 software and HDi systems with 15.0 software.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR

Full cause & step-by-step resolution