GE Healthcare · MRExcite22185372248008224715122471542253614199936
Scan stop caused by IRF I/O BD
Symptom
Scans Stop Intermittently With Error Code 2248008 "RFM EOS semaphore timeout error" See Long Description Scans stop in the middle of the scan with the following error logged. This happens about once per day. (PQR 13096684) 2248008 EXCITE15T NSP APS:DRF 250 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. Please change prescription or perform a TPS Reset. DRF ID: 1 FE ID: 1 segCount: 12 frameCount: 14
Error message
2218537 Scan stopped, Scan Stopped due to error. DRF reported error Error: 2218537 Host: DMR2-OC0 Process: scn File: MsaScanMgr.m Line: 3038 Scan stopped, Scan Stopped due to error : DRF reported error, see other error messages in error log for detail. Please perform a TPS Reset.. 2248008 RFM EOS semaphore timeout error. IRF did not send enough data to DRF. 2247151 IRF Driver Function Failed to write a control value to the RRF Mon May 12 14:20:12 2008 Error: 2247151 Host: UWMR1 Process: diagCoord File: Exciter.cpp Line: 816 Task ID: 0x7effdf0 Time: 1353 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF driver function Exciter::ex_set_clr_srf_theta failed to write a new control value to the RRF. Exciter=1, Written Value = 1, Returned Value = 0 If this occurs often you may have RRF chassis communication problem.-> 2247154 The IRF driver function Receiver rx_set_r1_atten failed to write a new control value to the RRF. Task ID: 0x7effdf0 Time: 1352 Object: APS Exception Class: Unknown Function: RF: RF Transmit The IRF driver function Receiver::rx_set_r1_atten failed to write a new control value to the RRF. Channel=4, Written Value = 0, Returned Value = -12 If th
Applies to
- Models
- Excite
- Product
- 3T Excite HDx 14.0 Software. The software that this solution is tied to is indicated by the Software concept below.
- Component
- 2274196 IRF I/O Board See Long Description. Per FEMC part information, Description of part number 2274196 is an INTERFACE and REMOTE RF TRANSITION BOARD. Part Comments as of 2/8/2008 are as follows; FLS - INTERFACE & REMOTE RF TRANSITION ADDED PER RSPL. USED ON PHOENIX MGD. MSA8429 05/14/02 ROUTE TO MR FOR CONSUMABLE COLLECTION PER B. HAHLEN, 7-11-02, A. MASCH THIS PART WAS ADDED AS PART OFTHE EXCITE NPI FOR MR. 9/01/02 MS24445 BH FLS - FE ASSIST 08OCT2002 (MS2045) ADDED TO MR DISTRIBUTION LIST FOR PARTS TRACKING.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.