Pular para o conteúdo principal
GE Healthcare · MRExcite22491292223817

Defective Patient Handling Power Supply Module PHPS.

Symptom

SRI Communication Fails

Error message

2249129 SRI Communication Error. A reset was not detected when requested to the SRI.

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Component
5215012 MR Patient Handling Power Supply see long description 5215012 MR Patient handling power supply for assembly for MR excite 3. This is identical to the 5109368 assembly but has power supplies with a modification to allow operation upto a 200 Gauss magnetic field.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues