Asset Scan Wont Run Due To Different Scan Range Than Asset Cal Scan.
Symptom
Will Not scan ASSET The only sequence that fails in this system is when the customer scans with the ASSET option.
Error message
2227990 No valid ASSET calibration was found. 2227990 File: APS:ed_rdb_hdr_check Line: 1 No valid ASSET calibration was found. An ASSET scan must have the same coil, position, entry, and landmark as the most recent ASSET calibration. 2218624 Failed to execute the next event. Error: 2218624 Host: lx-mr Process: scn File: StateMachine.c Line: 200 Failed to execute the next event. 2218528 Reference Scan Has Stopped Wed Nov 29 08:08:15 2006 Error: 2218528 Host: mr3c Process: scn File: MsaScanOpsScrn.m Line: 1622 Reference Scan has stopped. error please try again
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.