GE Healthcare · MRDiscoveryExcite22740332270004
Multicoil Bias Faults Due to Coil.
Symptom
Unable to Scan With Multi Coil Bias Faults Unable to scan with Multi Coil Bias Faults
Error message
2274033 Coil Fault Short Circuit A multicoil bias fault was detected while scanning Error: 2274033 Host: CMR3T Process: NSP File: SCP:RfHubCanHWO::RfBias Line: 3952 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: fse-xl Coil: HD Cardiac Scan: 3791/5 Coil Fault - Short Circuit A multicoil bias fault was detected while scanning. Connector: Port A First channel: 1 of 8, counting from 1 2270004 The VRF detected that MR Signal is too high (overrange) at channel: 0. see long description Mon May 18 14:10:08 2009 Error: 2270004 Host: signa2 Process: LxMxConsumer File: VrfTrace.cpp Line: 53 Server Name: ADS The VRF detected that MR Signal is too high (overrange) at channel: 0. The VRF is located at ICN1(ICN2), if ADS(ADS1) reports this error.
Applies to
- Models
- Discovery, Excite
- Product
- Discovery MR750. Also called DVMR. This product is compatible with 3.0T 20.0, 20.1_IB1, IB2, 21.0 and DV22.0 software only
- Software
- 20.1_IB1 and IB2 This release of software is applicable to Discovery MR450 and Discovery MR750 systems. See Long Description 20.1_IB1 is the unified Discovery MR450 and Discovery MR750 software release. Known internally to engineering as "DVIB". This release of software is applicable to Discovery MR450 and Discovery MR750 systems. 20.1_IB2 is the upgrade (bug fixes) for IB1.
- Component
- SecondOpinion 2414383 3T NV NO CABLE 3.0T HD Neurovascular Array 2414386 (with cable) and 2414383 (with out cable). Cable FRU# 2414385. See comments in resolution concept for second opinion requirements before ordering this part
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.