GE Healthcare · MRExcite2248349
Unable to Boot Applications Software due to Software Corruption Caused by Repeated Power Brownouts.
Symptom
Applications Software Will Not Boot.
Error message
2248349 The file MRconfig.cfg contains a missing or bad value. See Long Description This error will include one of many different data values that can be corrupted. Some examples are: begTravel boreTempHyst boreTempLev1 boreTempLev2 boreTempOffset digiTuningBoard endTravel fieldStrength gradientAmp gradProcessorType isoVectorZ lineFreq mCoilTRdriver PowerMonType realTimeMon RFampType rrfCalibFlag specRFamp spectroRFampType usssType
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR