Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova810020

Database Inconsistency Failure

Symptom

Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited. Database Inconsistency Failure: Continue and call service for repair is displayed on the image monitor. This issue may occur during "Start" of an exam or during an exam. It occures when a file corruption has occurred in the DL database. Database inconsistency failure and a Debug text of Check And Repair Requested at Browser initialization. When the customer boots the system, the image monitor displays Database inconsistency failure: Continue and call service for repair. This is a warning to the customer that file corruption has occurredon the image disk and a corrupt sequence may not network or display properly. Xray is not inhibited.

Error message

error Code = 810020, error Text = Database inconsistency failure, debug Text += CONSISTENCY ERROR: The number of frames stored in the DataBase and on the IP are inconsistent, recovery Class = WR, major Function /= XRDataMgr System impact: Database inconsistency. Displayed message : "Database inconsistency failure: continue and call service for repair".

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required