Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova10502635026310501031050042

INX100, Lost All Gantry axis's drive or NO movements

Symptom

INX100, Lost All Gantry axis's drive or NO movements The Gantry axis's will not drive. The Innova Gantry is locked and will not drive the L-arm, Pivot, Arc, or Lift. Emergency backout works, L-arm clutch releases. Gantry Problem: reset the system to recover is displayed on the monitor The Gantry axis' will not drive. The Innova Gantry is locked and will not drive the L-arm, Pivot, Arc, or Lift. Emergency backout works, L-arm clutch releases. Gantry Problem: reset the system to recover is displayed on the monitor

Error message

1050263 FRT LIFT AXIS SWITCH ENGAGED - Lift_Limit_Switch 1 0 - Limit switch X problem (prd:BB or BC) 50263 FRT LIFT AXIS SWITCH ENGAGED - Lift_Limit_Switch 1 0 - Limit switch X problem (prd:BB or BC) 1050263 FRT LIFT AXIS SWITCH ENGAGED WR GantryAxis No Motion Lift Limit Switch 0 0 - Limit switch 1 problem (prd: BB) 1050263 FRT LIFT AXIS SWITCH ENGAGED WR GantryAxis No Motion Lift_Limit_Switch 1 0 - Limit switch 2 problem (prd: BC) Z:/heroic/prd_task.c 638 This error will always be preceeded by error 1050103 and 1050042 1050263 FRT LIFT AXIS SWITCH ENGAGED - Lift_Limit_Switch 1 0 - Limit switch X problem (prd:BB or BC) 1050263 FRT LIFT AXIS SWITCH ENGAGED WR GantryAxis No Motion Lift Limit Switch 0 0 - Limit switch 1 problem (prd: BB) 1050263 FRT LIFT AXIS SWITCH ENGAGED WR GantryAxis No Motion Lift_Limit_Switch 1 0 - Limit switch 2 problem (prd: BC) Z:/heroic/prd_task.c 638 This error will always be preceeded by error 1050103 and 1050042 1050103 Pivot AXIS POWER WR GantryAxis No Motion Pivot AXIS POWER 1 0 - Power input from K3 problem (prd: BD) 1050042 Larm AXIS POWER WR GantryAxis No Motion Larm AXIS POWER 1 0 - Power input from K3 problem (prd: BD)

Applies to

Models
Innova
Product
Innova 2000 Innova 2000 systems Innova 2000 systems Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Positioner Gantry Positioner Gantry Positioner Gantry
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.