Pular para o conteúdo principal
GE Healthcare · MR

HFO cable run 830 (4 leg cable) is different from regular LCC run 830

Symptom

HFO cable run 830 (4 leg cable) is different from regular LCC run 830

Applies to

Product
Magnet HFO. Software Version Concept should NOT be selected for Magnet Products Serial # (Txxx) Field strength: 0.7 T He Vessel Pressure >0.9 psig but <6.25 psig Shield Temp for Cryo-cooler coldhead < 40 K Temp for Shield-cooler Coldhead < 35 K Cryocooler 2nd stage Temp > 3.8K but < 4.5K Recondenser Temp > 4.2K but < 4.6K Delta Temp (Coldhead - Recondensor) < 0 but < 0.3K Min. LHe Fill Level 75% of total vessel Min. LHe Ramp Level 100% Helium Volume 340 Liters Burst Disk 2253802-2 Baffle Assy 2226764
Component
2279237 HFO cable run 830 part Number The HFO 4 leg cable run 830 is part number 2279237, part of HFO SHEM cable kit 2350824. The HFO 4 leg cable run 830 is part number 2279237, part of HFO SHEM cable kit 2350824 2350824 HFO SHEM Cable kit
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues