Pular para o conteúdo principal
GE Healthcare · MRExcite222466622247252224580

Prescan Failure due to 8Ch CTL Coil Cable.

Symptom

Auto Prescan Failure

Error message

2224666 Auto Prescan failed, signal too small. Please try Manual Prescan. Fri Jan 23 14:39:55 2004 2224666 lx-mr psc PscMainControl.cc 1271 Auto Prescan failed, signal too small. Please try Manual Prescan.

2224725 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan) See Long Description Error: 2224725 Host: hopgemr1a Process: NSP File: APS:aps_frame_range_check Line: 10 Low signal detected. Please check the signal level for receiver X (or receiver X while using Manual Prescan).

2224580 Auto Prescan failed. Signal was too small to process - use Manual Prescan 2224580 Host: hopgemr1a Process: NSP File: APS:aps_cf Line: 11 Auto Prescan failed. Signal was too small to process - use Manual Prescan

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.1 For product 1.5T Excite only. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Component
2377426-7 USAI 8CH CTL COIL CABLE W/COIL ID 2377426-7USAI 8CH CTL COIL CABLE W/COIL ID
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.