GE Healthcare · X-ray VascularAdvantxInnova360206836020513602010

Chiller Warning Cooling System Failure,

Symptom

INX100: Chiller Warning Cooling System Failure,: "Thermal Issue, Xray aborted in 3 mins. Secure patient and call service. Chiller Warning: Cooling System Failure, 3602068 :Innova : Monitor displays "Thermal Issue, Xray aborted in 3 mins. Secure patient and call service. Chiller not working Error 3602051. The Chiller may illumate the red faucet icon, temperature icon, or display an error indication on the control panel.

Error message

error Code = 3602068, error Text = Jedi: Chiller warning followed within 3 minutes by error Code = 3602051, error Text = Jedi: Chiller not working System impact: X-ray unavailable. Displayed message : "Cooling Failure; Shut System down and call Service"error Code = 3602068, error Text = Jedi: Chiller warning , debug Text += jedi-error=812h, chiller problem, recovery Class = WR, major Function /= XRTechMgr followed within 3 minutes by error Code = 3602051, error Text = Jedi: Chiller not working, debug Text += jedi-error=902h, tube cooling supply error, recovery Class = IH, major Function /= XRTechMgr System impact: X-ray unavailable. Displayed message : "Cooling Failure; Shut System down and call Service" error Code = 3602068, error Text = Jedi: Chiller warning , debug Text += jedi-error=812h, chiller problem, recovery Class = WR, major Function /= XRTechMgr followed within 3 minutes by error Code = 3602051, error Text = Jedi: Chiller not working, debug Text += jedi-error=902h, tube cooling supply error, recovery Class = IH, major Function /= XRTechMgr. Here is a different pattern on site: system shutdowned during the night. There comes a thunderstorm that sithced OFF all room equip

Applies to

Models
Advantx, Innova
Product
Advantx E Monoplane Advantx E Systems advantx e, vamp Advantx E generator Systems Advantx E Biplane Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Advantx E biplane systems: LCLP+ (22 cm II), LCN+ (32 cm II), LCALP+ (40 cm frontal plane, 32 cm lateral plane) Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Coolix XR Tube Chiller Coolix XR Tube Chiller Coolix XR Tube Chiller
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.