Pular para o conteúdo principal
GE Healthcare · MRExciteDiscoveryOptima2247814

Cant Start A New Patient Due To Diagnostic Communications (DC) Error When Reading Socket Connection

Symptom

Unable to Start a New Patient Entry. This site had crashed numerous times as a result of a severe storm. After a successful reboot, and TPS RESET, the new patient entry was not responding and the system was reporting messages of Startup of Application failed, Reset TPS and MDS Link failure appears to be broken between SCP and the Gradient.

Error message

2247814 Diagnostic Communications (DC) error. Error when reading socket connection 2247814 Host: GEMS Process: diagCoord File: procDcMsg.c Line: 521 Diagnostic Communications (DC) error. Error when reading socket connection. An automatic attempt will be made to recover. File descriptor: 12 Object: Host Diag Executive Error code: -1-

Applies to

Models
Excite, Discovery, Optima
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR