Pular para o conteúdo principal
GE Healthcare · MRExcite2248014

Cannot scan thick FIESTA slabs back-to-back, error indicates insufficient memory.

Symptom

Cannot scan thick FIESTA slabs back-to-back, error indicates insufficient memory. See Long Description Can't start scan on Vascular Time-of-Flight SPGR using NVARRAY coil, 56 locations prescribed. Prescan works ok, then Prep-for-Scan fails with error message "not enough memory".

Error message

2248014 Allocate DPR memory error 2248014 - Allocate DPR memory error. Possible reason : too many printf or log msg caused APS missed EOS interrupt. Software error. Please change prescription or perform a TPS reset. DRF ID : 4 FE ID : 0 arg0 : 1 arg1 : 0

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required