Low BD4 HV Bias in Body and Surface Modes due to Driver Module

Symptom

Unable To Scan With Driver Module Errors Unable To Scan With Driver Module Errors.

Error message

2252312 - DD Driver Fault Detected - High Voltage Fault. Error: 2252312 Host: mr3t Process: NSP File: SCP:DCB Line: 1884 DD Driver Fault Detected - High Voltage Fault. Source of Fault is BD2 (Dynamic Disable Boards at Front of Body Coil). Value = 213 V Limit: 425 V->

2252312 -DD Driver Fault Detected - High Voltage Fault. Source of Fault is BDX (Dynamic Disable Boards at Back of Body Coil) Error: 2252312 Host: WC MR02 Process: NSP File: SCP:DCB Line: 2051 Exception Class: Unknown Severity: Unknown Function: RF: RF Bias PSD: Vasc TOF SPGR Coil: HDNV Array Scan: 8866/5 DD Driver Fault Detected - High Voltage Fault. Source of Fault is BDX (Dynamic Disable Boards at Back of Body Coil). Value = 0 V Limit: 300 V-

Applies to

Models
Signa, Excite
Product
3.0T Signa HDxT For 3.0T Signa HDxT systems with 15.0 software only. For 3.0T Signa HDxT systems with 15.0 software only.
Software
15.0 For Product Signa HDxT and HDi only. -- NOT to be used with HDe. -- For Product Signa HDxT and HDi only. -- NOT to be used with HDe. --
Component
SecondOpinion Excite HD & HDx Driver Module 5110612-2 See comments in resolution concept for second opinion requirements before ordering this part
Area
MR

Full cause & step-by-step resolution