Pular para o conteúdo principal
GE Healthcare · X-ray VascularInnova12001131200108

Innova System boot failure Any defect that prevents from getting the first acquisition after the system reset.

Innova System boot failure

Symptom

Innova System boot failure Any defect that prevents from getting the first acquisition after the system reset. Customer needs to boot again to recover. When Innova system has a actor issue.

Error message

error Code = 1200113, 1200108 error Text = ERR_DET_POWER_SUPPLY_BOOT_WARNING, debug Text += [DemSupervisor::_PWSStatusDebugMessage] Statuses are bad : (PG1B : OK) (PG2A : BAD) (PG2B : OK) (PG3 : OK) (PG1A : BAD)~ followed by error Code = 1200108 System impact: System boot failure. Displayed message : "Xray not available - Reset system; if problem persists, call Service". 1200108: Root Cause: Component(s) - Frontal Detector Power Supply This error is due to bad status sent by Frontal Detector Power Supply to Frontal RTAC Gitane. Recommended Action: Run RTAC subsystem//Front Detector Power Supply diagnostics 1200113: Root Cause: Component(s) - Frontal Detector Power Supply Warning This error is due to an unsucessful Detector Power supply boot attempt for frontal plane. Recommended Action: If the error is NOT repeated and NOT followed by a DetPS Boot Failure error, no Service recommended action

Applies to

Models
Innova
Product
Innova X1 X1 Biplane Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova X1 X1 Bi-plane Innova 2121 or 3131 Bi-plane System Innova 2000 Innova 2000 systems Innova 2000 systems Innova X100 Monoplane Innova X100 Monoplane from product name Innova 2100, 3100 and 4100 Innova X100 Monoplane Innova 2100, 3100 and 4100
Component
Detector and Detector Power Supply Detector PS 2302361-2 for 21 cm detector, 2302361-3 (alt # 2300475-2) for 31/41 detector Detector PS 2302361-2 for 21 cm detector, 2302361-3 (alt # 2300475-2) for 31/41 detector
Area
X-ray Vascular

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray Vascular issue are available to FSELIB subscribers — re-authored, web-first, searchable.