Pular para o conteúdo principal
GE Healthcare · MRExcite2250080

Scan Pauses with Gradient Errors Due to STIF board.

Symptom

Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.

Error message

2250080 X Axis Gradient Data Overrange Occurred. 2250080 MD-DMR-MRGE2 NSP SCP:Grad 3253 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Fiesta Coil: 8Ch Body Upper Scan: 4/15 X Axis Gradient Data Overrange Occurred.->

Applies to

Models
Excite
Product
Excite HD For 12.0 1.5T Systems only.
Software
12.0 Excite HD and 3T Excite HD products only
Component
2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues