GE Healthcare · MRExcite2250080
Scan Pauses with Gradient Errors Due to STIF board.
Symptom
Scan Pauses with Gradient Errors and Continues after start Scan Pressed then Images Show Bright Line Artifact System scans and pauses with Gradient errors. Images had a definite gradient collapsed.
Error message
2250080 X Axis Gradient Data Overrange Occurred. 2250080 MD-DMR-MRGE2 NSP SCP:Grad 3253 Exception Class: Unknown Severity: Unknown Function: Gradient: N/A PSD: Fiesta Coil: 8Ch Body Upper Scan: 4/15 X Axis Gradient Data Overrange Occurred.->
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2280952-3 STIF Board See Long Description SRF TRF I/O or STIF 2280952 crosses up in FEMC to 2280952-3. Part Description is "STIF Assembly".
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.