Symptom

Intermittent rotated images - may have no errors in the log. See Long Description for details Intermittent rotated images - may have no errors in the log.. Rotated images in all of the helical series. Rotated Image two.

Error message

Error: 260107214 Gantry Failed to Stop at Position XX degrees within XX.X seconds See Long Description for complete error Thu Sep 22 13:19:47 2005 Error: 260107214 Host: tgp Process: tgp_subsys File: amc_mgr.c Line: 3687 Thu Sep 22 13:19:47 2005 Host : Table Gantry Processor Ermes # : 260107214 Exception Class : Abort Severity : Primary File : amc_mgr.c Line# : 3687 Function : AXIAL CONTROL : Axial Control Scan Type : Scout Scan: 279/1/1 Gantry Failed to Stop at Position 90 degrees within 28.000 seconds. (AMC state:7 Cmd:24 CurPos:91.0) Retry Rx to verify probable Axial Drive or Gantry Balance or Brake or Encoder Hardware.

Applies to

Models
BrightSpeed, VCT
Product
BrightSpeed Select, Delight Select or MERI Select Elite-16, Edge-8, Excel-4 HP_S_G16_G_HLT HP_S_G8_G_HLT HP_S_G4_G_HLT BrightSpeed Select or Delight Select (Hivalue Lite) 4/8/16 Slice Scanner BrightSpeed Delight Select Systems started to ship in first Quarter or 2009 with AIO (All in One) Console. AIO console uses HP8400 and NO DARC or IG's. Recon time is 6 FPS. If customer orders Fluoro option system requires one IG. The 16 FPS Upgrade in Not Available for Select systems. BrightSpeed Meri Select Systems ( Pancake Das ) started to ship in 3rd quarter of 2010, 4/8/16 Slice Scanner HiValue 16 Lite −−> BrightSpeed Elite Select HiValue 8 Lite−−> BrightSpeed Edge Select HiValue 4 Lite−−> BrightSpeed Excel Select BrightSpeed Select Scope: - 3.5MHu Solarix tube - JEDI 60 dc Lite - Modified NP Table Key Features: - 16x0.625mm(16x1.25) for Lite 16 - 8x1.25/8x2.5mm for Lite 4/8 - Rotation Speed: 0.8 s - Gantry: 2040mm width, 70cm Bore - Table: 1337@ISO-140mm scannable range - Recon: 3fps (w/o IG @ 2G Memory(4/8), 4G Memory(16)), optional: 6fps(w/ 1IG ) - Power: 42KW JEDI power See the attached Service Intro Planner for details Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP, VCT, and HD750. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) See LD for details 5311936 Axial Encoder Assembly, Hub-Mounted Wheel. Used on H2, HP and VCT. (Replaces 5182284-2 /5182284 / 2369747-2 / 2369747 / 2328307) Part #: 5311936 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: Shipping collector for Axial Encoder Assembly. FRU: Yes Status: Orderable Comments: Per ECR 2058601, 212008898, 10/8/08
Area
CT

Full cause & step-by-step resolution