CWA Fails TPS Reset when booting up to Applications or when Resetting TPS.

Symptom

Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.

Error message

2270016 ADS1 VRF did not receive correct device ID from RRX module in scan room. RRX does not exist or is not connected to VRF. The VRF is located at ICN1(ICN2), Fri Feb 27 07:06:54 2009 Error: 2270016 Host: mr1 heri Process: LxMxConsumer File: VrfTrace.cpp Line: 80 Server Name: ADS1 VRF did not receive correct device ID from RRX module in scan room. RRX does not exist or is not connected to VRF. The VRF is located at ICN1(ICN2), if ADS(ADS1) reports this error. VRF ID: 0 Device ID: 2-

Applies to

Models
Optima, HiSpeed, Excite
Product
Optima MR450W Wide bore magnet. Not Discovery750 or Discovery450.
Software
DV22.0 For the Discovery MR750, MR450 or Optima MR450w products only. For the Discovery MR750, MR450 or Optima MR450w products only.
Area
MR