Application Software does not complete the boot process with TPS Download failure Due to VRE Configuration.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2223852 The APS board reported a communication manager (CM) initialization error. See Long Description Please check the error log for a more detailed error message from the board. Extended Error: MCP820 boards can be identified by the word MCP820 printed on the board extraction lever 2249436 Communication problem between APS and VRE. See long description Mon Jun 4 07:37:51 2007 Error: 2249436 Host: lx-mr Process: diagCoord File: cm_init.c Line: 293 Task ID: 0x6d89800 Time: 1807 Object: APS Exception Class: Unknown Function: Software/Firmware: Host Communications Communication problem between APS and VRE. The APS could not notify the VRE that communication is established.-> Extended Error: ERROR: EM_OS_CM_INIT_RPC_VRE Communication problem between APS and VRE. The APS could not notify the VRE that communication is established. Extended Error: This could be an indication of a hardware problem with the APS or VRE boards or an ethernet cabling problem on the MGD subnet. Verify the ethernet cabling on the MGD subnet, reboot VRE, and perform a TPS Reset. Run diagnostics if the problem recurs. 2223880 Reset TPS was not successful. Board Failure Status APS Fail, SCP: Pass, AGP Pass, VRE
Applies to
- Models
- Excite
- Product
- Excite HDe. For Hino HDe 14.0 and HDe 15.0 Software Systems only
- Software
- HDe 15.0 For Hino Excite HDe 15.0 systems only.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- TPS/MGD fails to download or reset. mgd_term windows fail to mount file system.
- The Carriage Fails to Advance to Latch to the Table and to Retract When it is Raised, Undocked or Lowered.
- Unable to Load Applications Software during restoreinfo Due to Corrupt saveinfo Disk.
- CWA ID100315. Auto and Manual Prescan Fails.
- Ghosted or Smeared Head Images Using Head Coil due to Head TR Assembly.
- A Copper Coolant Line or Pipe Internal to the MCS Cabinet Chiller Burst due to Water Remaining in Unit Freezing During Shipment.