Pular para o conteúdo principal
GE Healthcare · MRSignaExcite22255402225542

2225540 Dynamic Disable Bias Fault on Front Drivers Due to the SSM.

Symptom

Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.

Error message

2225540 Dynamic disable fault, Front of Coil short. Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM. 2225540 Dynamic disable fault, Front of Coil short Dynamic Disable bias fault on front drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on FRONT of BRM.

2225542 Dynamic Disable bias fault on back drivers Dynamic Disable bias fault on back drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on back of BRM.

Applies to

Models
Signa, Excite
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
SecondOpinion System Support Module 2160200-43 See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH See comments in resolution concept for second opinion requirements before ordering this part. FE replaced SSM which is now a FRU Part #: 2160200-43 Part Type: GP REPAIRABLES Part Weight: 48.50 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ERBTEC #550020 SYSTEM SUPPORT MODULE FRU: Yes Status: Orderable ICV/Transfer Price: 5684.00 USD /Each List Price: 36215.22 USD /EACH
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.